Utility of multi‐frequency bioelectrical impedance compared to deuterium dilution for assessment of total body water

Principle/Accuracy
Author
Lindsey J. ANDERSON et al.
Used Parameters : TBW

Publication Year : 2015

Country : USA

Journal : Nutrition & Dietetics

Aim Multi‐frequency bioelectrical impedance analysis (MF‐BIA) may provide a quick, cost‐efficient method of total body water (TBW) assessment. We compared TBW estimates from the B iospace I n B ody 520, 720, and S 10 MF‐BIA devices to TBW estimates from isotopic deuterium dilution (D 2 O). Methods TBW was measured in 25 men and 25 women, 18–49 years, 73.6±15.4 kg on each of the three devices and from D 2 O analysis. Results All MF‐BIA devices significantly overestimated TBW compared to D 2 O TBW for men (4.7–5.4 kg …
Full text